Page 1 Page 2 Page 3 Page 4 Page 5 Page 6 Page 7 Page 8 Page 9 Page 10 Page 11 Page 12 Page 13 Page 14 Page 15 Page 16 Page 17 Page 18 Page 19 Page 20sales@goSTI.cc 7 · ·Color Touch Screen Graphical User Interface · ·Real-Time Graphing Mode · ·Bi-Directional Flow Measurement · · Data Storage Using SD Flash Card and Internal Memory · ·Rechargeable Battery Plus AC Operation · ·Flow with Peak & Minimum Flow · ·Volume (Inhaled and Exhaled) · ·Minute Volume · ·Low Pressure (Differential) · ·Peak & PEEP Pressure · ·Mean Airway Pressure · ·Barometric & High Pressure · ·Inspiratory/Expiratory Time · ·I:E Ratio · ·Respiratory Rate · ·Gas Temperature · ·Oxygen Concentration with optional 4073 Kit Certifier FA Plus 4080 Ventilator Test System Certifier FA PLUS Gas/Mode Range Accuracy** Flow—High Flow Air, O2 -200 to +300 slpm* ±2% or ±0.075 slpm Air/O2 Mixtures 0 to 300 slpm ±4% or ±0.1 slpm N2 -200 to +300 slpm ±3% or ±0.1 slpm CO2 -40 to +40 slpm ±3% or ±0.1 slpm Flow—Low Flow Air, O2 0.01 to 20 slpm ±2% or ±0.01 slpm N2, CO2 0.01 to 20 slpm ±3% or ±0.01 slpm N2O 0.01 to 20 slpm ±4% or ±0.025 slpm Volume-High Flow-Inhaled Air, O2 0.01 to 10 liters STP ±2% Plus 0.02 liters Air/O2 Mixtures 0.01 to 10 liters STP ±4% Plus 0.02 liters Volume-Low Flow-Inhaled Air, O2 0.01 to 10 liters STP ±2% or ±0.01 liters N2O 0.01 to 10 liters STP ±4% or ±0.01 liters Minute Volume-High Flow 0.01 to 100 liters STP ±3% Minute Volume-Low Flow 0 to 10 liters STP ±3% Respiratory Times Ti, Tip, Te 0.04 to 30 seconds ±2% or ±0.01 seconds I:E Ratios I:E, I:Eip 1:100 to 100:1 ±4% Respiratory Rate f 1 to 1500 breaths per minute ±2% or 0.1 bpm Low Pressure All -25 to +150 cm H2O ±0.5% or ±0.15 cm H2O High Pressure -10 to +150 PSI (-0.7 to 10 bar) ±1% or 0.1 PSI (7 mbar) Barometric Pressure 7 to 23 PSI (500 to 1600 mbar) ±0.16 PSI (11 mbar) Oxygen Concentration 21% to 100% 2% of concentration *slpm = Standard Liters per Minute ** Accuracy stated as a percent of reading at TSI standard gas conditions. Certifier FA Plus is the full-feature system capable of testing virtually all models of ventilators: adult, pediatric, anesthesia, neonatal and high-frequency. PRECISION MEASUREMENT TOOLS WITH UNSURPASSED ACCURACY AND RELIABILITY